Fluke MET/CAL Procedure ============================================================================= INSTRUMENT: HP 54602A: (1 year) CAL VER /5800 DATE: 27-Aug-98 AUTHOR: Fluke Corporation REVISION: $Revision: 1.1 $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 28 NUMBER OF LINES: 354 CONFIGURATION: Fluke 5800A ============================================================================= # # Source: Hewlett Packard 54600-Series User and Service Guide, June 1992 # # Compatibility: MET/CAL 5.0 or later # # Subprocedures: None # # Required Files: None # # System Specifications: # TUR calculation is based on specification interval of the accuracy file. # The default 5800A accuracy file contains 90 day specs. # # Fluke makes no warranty, expressed or implied, as to the fitness # or suitability of this procedure in the customer's application. # STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- R N J F W 1.002 ASK+ D 1.003 HEAD PRELIMINARY INSTRUCTIONS 1.004 DISP Warm-up time: 30 min. - REQUIRED 1.004 DISP Temperature : 20 - 30 degC 1.004 DISP Humidity : <80% 1.005 HEAD [32] WARNING! 1.006 DISP 1.006 DISP DURING THE FOLLOWING PROCEDURE, HIGH VOLTAGES THAT COULD 1.006 DISP CAUSE SERIOUS INJURY OR EVEN DEATH ARE PRESENT! 1.006 DISP 1.006 DISP USE CAUTION TO PREVENT INJURY! 1.007 DISP This procedure requires the following... 1.007 DISP 1.007 DISP Digital Multimeter; 0.1% mV resolution, 0.01% accuracy 1.008 HEAD {DC CALIBRATOR} 1.009 DISP Connect the multimeter to the rear panel DC Calibrator. 1.010 DISP Press PRINT/UTILITY 1.010 DISP 1.010 DISP Press the Self Test softkey, then press the DAC softkey. 1.011 HEAD DC CALIBRATOR: {Low Level} 1.012 EVAL Is the {DC Calibrator output 0.00 VDC +/-500 uV}? 2.001 HEAD DC CALIBRATOR: {High Level} 2.002 DISP Press any key to continue the test. 2.003 EVAL Is the {DC Calibrator output 5.00 VDC +/-10 mV}? 3.001 MATH M[100] = 1 3.002 LABEL VOLTAGE_MEASUREMENT_ACCURACY 3.003 HEAD VOLTAGE MEASUREMENT ACCURACY CH [M100] 3.004 RSLT =CH [M100] 3.005 DISP MAKE THE FOLLOWING CONNECTION... 3.005 DISP 3.005 DISP 5800A SCOPE OUTPUT CH 1 ----------- CH [M100] INPUT 3.006 JMPL CH2_VOLTAGE_MEASUREMENT M[100] == 2 3.007 DISP Press SETUP, then press the Default Setup softkey. 3.007 DISP 3.007 DISP Press VOLTAGE, then press the V avg softkey. 3.008 DISP Set the Volts/Div to 5 V/div. 3.009 DISP Adjust the CH [M100] Position knob to place the baseline 3.009 DISP near (but not at) the bottom of the display. 3.010 DISP Press CURSORS, then press the V1 softkey. 3.011 5800 0V S6 S 3.012 DISP Using the cursors knob, set V1 cursor on the baseline. 3.012 DISP 3.012 DISP If you are in an electrically noisy environment, 3.012 DISP it can help to place a shorting cap on the input BNC 3.012 DISP connector when positioning V1. 3.013 HEAD VOLTAGE MEASUREMENT ACCURACY CH [M100]: {5 V/div} 3.014 5800 35V S6 S 3.015 DISP Press the V2 softkey, then position the V2 cursor 3.015 DISP to the baseline. 3.016 EVAL Is {Delta V readout between 34.315 and 35.685 V}? 4.001 HEAD VOLTAGE MEASUREMENT ACCURACY CH [M100]: {2 V/div} 4.002 DISP Set the Volts/Div to 2 V/div. 4.003 5800 0V S6 S 4.004 DISP Using the cursors knob, set V1 cursor on the baseline. 4.004 DISP 4.005 5800 14V S6 S 4.006 DISP Press the V2 softkey, then position the V2 cursor 4.006 DISP to the baseline. 4.007 EVAL Is {Delta V readout between 13.726 and 14.274 V}? 5.001 HEAD VOLTAGE MEASUREMENT ACCURACY CH [M100]: {1 V/div} 5.002 DISP Set the Volts/Div to 1 V/div. 5.003 5800 0V S6 S 5.004 DISP Using the cursors knob, set V1 cursor on the baseline. 5.004 DISP 5.005 5800 7V S6 S 5.006 DISP Press the V2 softkey, then position the V2 cursor 5.006 DISP to the baseline. 5.007 EVAL Is {Delta V readout between 6.863 and 7.137 V}? 6.001 HEAD VOLTAGE MEASUREMENT ACCURACY CH [M100]: {0.5 V/div} 6.002 DISP Set the Volts/Div to 0.5 V/div. 6.003 5800 0V S6 S 6.004 DISP Using the cursors knob, set V1 cursor on the baseline. 6.004 DISP 6.005 5800 3.5V S6 S 6.006 DISP Press the V2 softkey, then position the V2 cursor 6.006 DISP to the baseline. 6.007 EVAL Is {Delta V readout between 3.4315 and 3.5685 V}? 7.001 HEAD VOLTAGE MEASUREMENT ACCURACY CH [M100]: {0.2 V/div} 7.002 DISP Set the Volts/Div to 0.2 V/div. 7.003 5800 0V S6 S 7.004 DISP Using the cursors knob, set V1 cursor on the baseline. 7.004 DISP 7.005 5800 1.4V S6 S 7.006 DISP Press the V2 softkey, then position the V2 cursor 7.006 DISP to the baseline. 7.007 EVAL Is {Delta V readout between 1.3726 and 1.4274 V}? 8.001 HEAD VOLTAGE MEASUREMENT ACCURACY CH [M100]: {0.1 V/div} 8.002 DISP Set the Volts/Div to 0.1 V/div. 8.003 5800 0V S6 S 8.004 DISP Using the cursors knob, set V1 cursor on the baseline. 8.004 DISP 8.005 5800 700mV S6 S 8.006 DISP Press the V2 softkey, then position the V2 cursor 8.006 DISP to the baseline. 8.007 EVAL Is {Delta V readout between 686.3 and 713.7 mV}? 9.001 HEAD VOLTAGE MEASUREMENT ACCURACY CH [M100]: {50 mV/div} 9.002 DISP Set the Volts/Div to 50 mV/div. 9.003 5800 0V S6 S 9.004 DISP Using the cursors knob, set V1 cursor on the baseline. 9.004 DISP 9.005 5800 350mV S6 S 9.006 DISP Press the V2 softkey, then position the V2 cursor 9.006 DISP to the baseline. 9.007 EVAL Is {Delta V readout between 343.15 and 356.85 mV}? 10.001 HEAD VOLTAGE MEASUREMENT ACCURACY CH [M100]: {20 mV/div} 10.002 DISP Set the Volts/Div to 20 mV/div. 10.003 5800 0V S6 S 10.004 DISP Using the cursors knob, set V1 cursor on the baseline. 10.004 DISP 10.005 5800 140mV S6 S 10.006 DISP Press the V2 softkey, then position the V2 cursor 10.006 DISP to the baseline. 10.007 EVAL Is {Delta V readout between 137.26 and 142.74 mV}? 11.001 HEAD VOLTAGE MEASUREMENT ACCURACY CH [M100]: {10 mV/div} 11.002 DISP Set the Volts/Div to 10 mV/div. 11.003 5800 0V S6 S 11.004 DISP Using the cursors knob, set V1 cursor on the baseline. 11.004 DISP 11.005 5800 70mV S6 S 11.006 DISP Press the V2 softkey, then position the V2 cursor 11.006 DISP to the baseline. 11.007 EVAL Is {Delta V readout between 68.63 and 71.37 mV}? 12.001 HEAD VOLTAGE MEASUREMENT ACCURACY CH [M100]: {5 mV/div} 12.002 DISP Set the Volts/Div to 5 mV/div. 12.003 5800 0V S6 S 12.004 DISP Using the cursors knob, set V1 cursor on the baseline. 12.004 DISP 12.005 5800 35mV S6 S 12.006 DISP Press the V2 softkey, then position the V2 cursor 12.006 DISP to the baseline. 12.007 EVAL Is {Delta V readout between 34.155 and 35.845 mV}? 13.001 HEAD VOLTAGE MEASUREMENT ACCURACY CH [M100]: {2 mV/div} 13.002 DISP Set the Volts/Div to 2 mV/div. 13.003 5800 0V S6 S 13.004 DISP Using the cursors knob, set V1 cursor on the baseline. 13.004 DISP 13.005 5800 14mV S6 S 13.006 DISP Press the V2 softkey, then position the V2 cursor 13.006 DISP to the baseline. 13.007 EVAL Is {Delta V readout between 13.47 and 14.53 mV}? 14.001 MATH M[100] = M[100] + 1 14.002 JMPL VOLTAGE_MEASUREMENT_ACCURACY M[100] == 2 14.003 JMP 14.009 14.004 LABEL CH2_VOLTAGE_MEASUREMENT 14.005 5800 1.0V S6 S 14.006 DISP Press AUTOSCALE 14.006 DISP 14.007 5800 * S 14.008 JMP 3.007 14.009 LABEL CH4_VOLTAGE_MEASUREMENT_ACCURACY 14.010 RSLT =CH [M100] 14.011 HEAD VOLTAGE MEASUREMENT ACCURACY CH [M100] 14.012 DISP MAKE THE FOLLOWING CONNECTION... 14.012 DISP 14.012 DISP 5800A SCOPE OUTPUT CH 1 ----------- CH [M100] INPUT 14.013 5800 1V 1kH SI S6 S 14.014 DISP Press AUTOSCALE 14.014 DISP 14.015 5800 * S 14.016 DISP Press VOLTAGE, then press the V avg softkey. 14.017 DISP Set the Volts/Div to .5 V/div. 14.018 DISP Adjust the CH [M100] Position knob to place the baseline 14.018 DISP near (but not at) the bottom of the display. 14.019 5800 0V S6 S 14.020 DISP Press CURSORS, then press the V1 softkey. 14.020 DISP Using the cursors knob, set V1 cursor on the baseline. 14.021 HEAD VOLTAGE MEASUREMENT ACCURACY CH [M100]: {0.5 V/div} 14.022 5800 3.5V S6 S 14.023 DISP Press the V2 softkey, then position the V2 cursor 14.023 DISP to the baseline. 14.024 EVAL Is {Delta V readout between 3.4315 and 3.5685 V}? 15.001 HEAD VOLTAGE MEASUREMENT ACCURACY CH [M100]: {0.1 V/div} 15.002 DISP Set the Volts/Div to 0.1 V/div. 15.003 5800 0V S6 S 15.004 DISP Using the cursors knob, set V1 cursor on the baseline. 15.004 DISP 15.005 5800 700mV S6 S 15.006 DISP Press the V2 softkey, then position the V2 cursor 15.006 DISP to the baseline. 15.007 EVAL Is {Delta V readout between 686.3 and 713.7 mV}? 16.001 MATH M[100] = M[100] + 1 16.002 JMPL CH4_VOLTAGE_MEASUREMENT_ACCURACY M[100] == 4 16.003 HEAD {BANDWIDTH} 16.004 MATH M[100] = 1 16.005 LABEL BANDWIDTH_CH1,2 16.006 RSLT =CH [M100] 16.007 DISP MAKE THE FOLLOWING CONNECTION... 16.007 DISP 16.007 DISP 5800A SCOPE OUTPUT CH 1 ------50 Ohm--CH [M100] INPUT 16.008 5800 800mVpp 250kH LS S6 S L 16.009 DISP Press AUTOSCALE. 16.010 MESS Slew the amplitude for 8 divisions. 16.011 5800 800mVpp 250kH LS S6 N L 16.012 DISP Change the Sweep Speed to 5 nS. 16.012 DISP 16.013 MESS Slew the frequency for 5.66 divisions. 16.014 5800 150MH -1000% mVpp LS S6 L 17.001 MATH M[100] = M[100] + 1 17.002 JMPL BANDWIDTH_CH1,2 M[100] == 2 17.003 LABEL BANDWIDTH_CH3,4 17.004 RSLT =CH [M100] 17.005 DISP MAKE THE FOLLOWING CONNECTION... 17.005 DISP 17.005 DISP 5800A SCOPE OUTPUT CH 1 ------50 Ohm--CH [M100] INPUT 17.006 5800 800mVpp 250kH LS S6 S L 17.007 DISP Press AUTOSCALE. 17.008 DISP Set the Volts/Div to 0.1 V/div. 17.009 MESS Slew the amplitude for 8 divisions. 17.010 5800 800mVpp 250kH LS S6 N L 17.011 MESS 17.012 DISP Change the Sweep Speed to 5 nS. 17.012 DISP 17.013 5800 250MH mVpp LS S6 S L 17.014 EVAL Is {the amplitude greater than 5.66 divisions}? 18.001 MESS 18.002 MATH M[100] = M[100] + 1 18.003 JMPL BANDWIDTH_CH3,4 M[100] == 4 18.004 ASK+ U 18.005 HEAD {HORIZONTAL DELTA T AND 1/DELTA T ACCURACY} 18.006 DISP MAKE THE FOLLOWING CONNECTION... 18.006 DISP 18.006 DISP 5800A SCOPE OUTPUT CH 1 ----------50 Ohm--CH 1 INPUT 18.007 HEAD HORIZONTAL DELTA T AND 1/DELTA T ACCURACY: {20 uS/div} 18.008 5800 0.1mT M1 S6 S L 18.009 DISP Press SETUP, then press the Default Setup softkey. 18.009 DISP 18.009 DISP Press AUTOSCALE. 18.010 DISP Set the time base to 20 uS/div. 18.011 DISP Adjust the trigger level to obtain a stable display. 18.011 DISP 18.012 DISP Press TIME, then press the Freq and Period softkeys. 18.013 EVAL Is {Frequency readout between 9.899 and 10.10 kHz}? 19.001 5800 0.1mT M1 S6 S L 19.002 EVAL Is {Period readout between 98.98 and 101.01 uS}? 20.001 HEAD HORIZONTAL DELTA T AND 1/DELTA T ACCURACY: {200 nS/div} 20.002 5800 1uT M1 S6 S L 20.003 DISP Set the time base to 200 nS/div. 20.004 DISP Adjust the trigger level to obtain a stable display. 20.004 DISP 20.005 EVAL Is {Frequency readout between 989.8 kHz and 1.0104 MHz}? 21.001 5800 1uT M1 S6 S L 21.002 EVAL Is {Period readout between 989.7 ns and 1.010 uS}? 22.001 HEAD HORIZONTAL DELTA T AND 1/DELTA T ACCURACY: {5 nS/div} 22.002 5800 10nT M4 S6 S L 22.003 DISP Set the time base to 5 nS/div. 22.004 DISP Set the Volts/Div to 0.5 V/div. 22.005 DISP Use the Position control to center the signal. 22.006 DISP Adjust the trigger level to obtain a stable display. 22.006 DISP 22.007 EVAL Is {Frequency readout between 97.078 and 103.1 MHz}? 23.001 5800 10nT M4 S6 S L 23.002 EVAL Is {Period readout between 9.699 and 10.30 nS}? 24.001 ASK- U 24.002 MATH M[100] = 1 24.003 LABEL TRIGGER_SENSITIVITY_CH2 24.004 HEAD CH [M100] INTERNAL TRIGGER SENSITIVITY 24.005 JMPL CH2_SETUP M[100] == 2 24.006 DISP Press SETUP, then press the Default Setup softkey. 24.007 HEAD CH [M100] TRIGGER SENSITIVITY: {25 MHz @ 0.35 div}? 24.008 5800 50mVpp 25MH LS S6 S L 24.009 DISP Press AUTOSCALE. 24.010 DISP Set the Volts/Div to 0.1 V/div. 24.011 MESS Slew the amplitude for 0.35 divisions. 24.012 5800 50mVpp 25MH LS S6 N L 24.013 MESS Adjust Trigger Level if needed for a triggered display. 24.013 MESS 24.014 EVAL Will the {display trigger}? 25.001 MESS 25.002 HEAD CH [M100] TRIGGER SENSITIVITY: {150 MHz @ 1.0 div}? 25.003 5800 100mVpp 150MH LS S6 S L 25.004 DISP Press AUTOSCALE. 25.005 DISP Set the Volts/Div to 0.1 V/div. 25.006 MESS Slew the amplitude for 1.0 division. 25.007 5800 100mVpp 150MH LS S6 N L 25.008 MESS Adjust Trigger Level if needed for a triggered display. 25.008 MESS 25.009 EVAL Will the {display trigger}? 26.001 MESS 26.002 MATH M[100] = M[100] + 1 26.003 JMPL TRIGGER_SENSITIVITY_CH2 M[100] == 2 26.004 JMP 26.008 26.005 LABEL CH2_SETUP 26.006 DISP MAKE THE FOLLOWING CONNECTION... 26.006 DISP 26.006 DISP 5800A SCOPE OUTPUT CH 1 ----------50 Ohm--CH 2 INPUT 26.007 JMP 24.006 26.008 LABEL INTERNAL_TRIGGER_SENSITIVITY_CH3,4 26.009 RSLT = CH [M100] 26.010 HEAD CH [M100] INTERNAL TRIGGER SENSITIVITY 26.011 DISP MAKE THE FOLLOWING CONNECTION... 26.011 DISP 26.011 DISP 5800A SCOPE OUTPUT CH 1 ------50 Ohm--CH [M100] INPUT 26.012 DISP Press SETUP, then press the Default Setup softkey. 26.013 DISP Select Channel [M100] 26.014 DISP Set the Volts/Div to 0.1 V/div. 26.015 HEAD CH [M100] TRIGGER SENSITIVITY: {25 MHz @ 0.35 div}? 26.016 5800 200mVpp 25MH LS S6 S L 26.017 DISP Press AUTOSCALE. 26.018 MESS Slew the amplitude for 0.35 divisions. 26.019 5800 50mVpp 25MH LS S6 N L 26.020 MESS Adjust Trigger Level if needed for a triggered display. 26.020 MESS 26.021 EVAL Will the {display trigger}? 27.001 MESS 27.002 HEAD CH [M100] TRIGGER SENSITIVITY: {250 MHz @ 1.0 div}? 27.003 5800 200mVpp 250MH LS S6 S L 27.004 DISP Press AUTOSCALE. 27.005 MESS Slew the amplitude for 1.0 division. 27.006 5800 100mVpp 250MH LS S6 N L 27.007 MESS Adjust Trigger Level if needed for a triggered display. 27.007 MESS 27.008 EVAL Will the {display trigger}? 28.001 MESS 28.002 MATH M[100] = M[100] + 1 28.003 JMPL INTERNAL_TRIGGER_SENSITIVITY_CH3,4 M[100] == 4 28.004 END